We check and adjust mass flow controllers, gauges, thermocouples, RF generators and lamp intensity against calibrated reference instruments, and record the values before and after.
A process recipe is only as good as the numbers the tool reports. On a tool that has run for twenty years, a mass flow controller can read a few percent off its true flow, a capacitance manometer can drift by several percent, and a thermocouple can age until the furnace runs hotter than its display says. Drift of this kind rarely triggers an alarm. It shows up as a slow change in etch rate, film thickness or line width, and as recipes that no longer transfer between tools.
Regular calibration of process equipment brings the reported values back in line with reality, and the as-found record tells you whether past results may have been affected.
The scope depends on the tool. Typical items in scope are:
For preventive maintenance customers we usually combine calibration with a scheduled visit, which keeps downtime to one window.
For most items the tolerance is the original specification; where your process needs a tighter limit, we use yours, and the record states which applies. Items found within tolerance are left alone, because adjusting a working instrument adds risk without benefit. Items outside tolerance are adjusted only where the hardware allows it. Some cannot be brought back, such as a mass flow controller with a contaminated sensor or a gauge head at the end of its life; those are flagged and quoted for replacement or rebuild rather than forced into range.
A calibration visit on a single tool typically takes from a few hours to a day, depending on how many items are in scope and whether the tool needs to reach process temperature. Furnace profiling takes longer because each zone has to stabilise.
We need:
You receive a calibration record per tool listing each item, its tolerance, the as-found and as-left values, the reference instrument used and its own calibration date. Items that were found out of tolerance are flagged with a note on which processes they may have affected, so you can judge whether earlier results need a second look. For mask aligners the record includes the intensity map; on platforms such as the SUSS MA6 this is also the starting point if you are considering a UV-LED source.
We list every calibrated item on the tool and agree which ones are in scope.
An engineer measures each item against a calibrated reference instrument and records the as-found value.
We adjust items that are out of tolerance and replace those that cannot be brought back.
We record the as-left values and give you a calibration record per tool.
200 mm (8 in) · Late 1990s–2010s
Each record lists the reference instrument used and its calibration date. If your quality system needs traceable or accredited calibration, tell us before we quote and we will say what we can document.
It depends on the gas, the process and how critical the flow is to your results. We recommend an interval per item based on its as-found drift.
Yes. Intensity and uniformity are measured across the exposure field with a calibrated UV intensity meter at the tool's wavelengths, and we adjust the lamp, optics or LED source to bring them back within specification.
We check metrology tools against reference samples and adjust them where the tool allows. Tell us the model and the reference samples you hold.
Reflectometers measure routine films quickly and cheaply. Ellipsometers handle very thin films, unknown materials and stacks. How to choose for your lab.
Daily, weekly, monthly and yearly checks that keep an RIE or ICP etcher stable, and the log entries that warn you before a run goes wrong.
Lamp house, stage, alignment optics and mask holder sizes decide what an older aligner can still do. A checklist for your inspection visit.
Tell us the tool and what you need. We scope the work and send a written offer.